The Coupled Column System
Unfortunately, samples can rarely be directly placed onto
the isotachophoretic instrument for analysis, as most
will invariably need some form of sample preparation. Sample preparation
techniques can vary widely in complexity from simple dilution to a complex
extraction followed by derivatization procedures. A
frequent problem involving sample preparation arises when the substances of
interest are included in mixtures containing high concentrations of other
materials. By employing a column
coupling procedure a large amount of sample can be used without the need for
sample pre-treatment such as extraction followed by concentration. The coupled column system that will be
described has a wide range of application as it can also be used to employ
other techniques in tandem with isotachophoresis such
as liquid chromatography or zone electrophoresis etc.
A diagram of a coupled column isotachophoresis
instrument is shown in figure 15. The coupled column apparatus consists of
three main parts, vis.,
1. A
pre-separation section
2. A
bifurcation section with a fitted detector
3. A
final separation section
In turn the bifurcation block has also three important
components, viz.,
a A
relatively wide bore hole (0.8 mm) in which the detector sensor(s) is mounted
and which is connected to the wide bore separation tube in the pre-separation
section.-
b
A narrow bore hole (0.2 mm) in line with the wide bore hole
(0.8 mm) to which the PTFE narrow bore
separation tube is connected.
C A rectangular hole (0.05mm x 1.0 mm) at right-angles to the other holes
which forms an electrical connection to the counter electrode compartment
applied during the pre-preparation.
![]()
![]()
![]()
![]()

Figure
15. A Coupled
Column Apparatus
A diagram of a commercially available isotachophoresis
instrument manufactured by Flowspek is shown in
figure 16.
The separation sections are in line with each other so that
zone mixing is minimised during trapping. To select the chosen band with
precision, the time interval between the detector sensing this band of interest
and the junction to the separation section can be calculated from the value of dV/dt of the constant current power supply in a leading
electrolyte/terminating electrolyte experiment. Thus, those zones requiring
further separation can be selected even if they are not migrating sequentially.

Figure
16. A Commercial Isotachophoresis Instrument
The choice of the bands of interest can be programmed so
that the selection is automatic and consequently highly reproducible. Using this
technique, the load that can be used for analysis can be increased by an order
of magnitude.