Quantitative Analysis
In figure 19, the same analysis is displayed. The linear
trace is produced by the potential detector but the widths of the steps (as
opposed to the step heights), provides the quantitative information the
magnitudes of which, are accurately provided by the differential trace.

Figure
19. The Analysis
of a Mixture of Two Substances, Perchlorate and p-Amino Benzoate that are Monitored by a Potentiometric
Detector.
The relationship between the zone lengths and the quantity
of ion species introduced into the system is linear assuming the current is
constant and well stabilized. For accurate quantitative results calibration
curves should be constructed covering the concentration ranger of interest and
example of which is given diagrammatically in figure 20.

Figure
20. A Calibration
Curve Obtained from a Set of Standard Isotachopherograms
If the mobility and the pKa
of an ion are known the calibration curve can be determined, An
internal standard can also be employed if so desired and can provide equally
accurate results. In general, however, it is better to work using calibration
curves for reliable and constant accuracy.