Instrumentation for Isotachophoresis
The Single Column System
In general an isotachophoretic
separation is carried out in a PTFE tube 0.2 to 0.4 mm internal diameter. In
most instruments the system is fitted with both conductivity and UV absorption
detection. A typical example of an isotachophoretic
instrument is shown in figure 14.

Figure
14. Basic Isotachophoretic Apparatus
The apparatus consists of two electrode compartments
carrying the leading and terminating electrolytes respectively in which are
inserted the anode and cathode, which in turn, are connected to a constant
current power supply. A PTFE tube, with inside and outside diameters of 0.2 and
0.3 mm respectively, has been found to be optimum for this type of separation.
Tubes of these dimensions have very small convective contributions and minimal
temperature affects. The sample is introduced by injection into the sample
block. As the zones are formed each one
will assume its own physical characteristics such as temperature, conductance,
pH, and potential gradient. In addition some zones may exhibit other
characteristics such as radioactivity, UV absorption optical rotation, and
fluorescence all of which, using appropriated detecting systems can aid in
sample identification. One of the first detection methods was thermal
measurement employing very thin constantan/copper thermocouples and although
the signal is small it is one of the more sensitive detection devices.
Potential gradient detection and conductivity detection is carried out using 25-micron
diameter electrodes. The electrodes are cemented into the detector block and
the sensor volume is a few nanolitres. The UV sources
is normally a low pressure mercury vapour lamp and light is guided by an
optical quality glass rod through the separation tube and then by another
optical glass rod system onto a suitable photoelectric cell. The different
parts of the apparatus are labelled in figure 14.
The Coupled Column
System
Unfortunately, samples can rarely be directly placed onto
the isotachophoretic instrument for analysis, as most
will invariably need some form of sample preparation. Sample preparation
techniques can vary widely in complexity from simple dilution to a complex
extraction followed by derivatization procedures. A
frequent problem involving sample preparation arises when the substances of
interest are included in mixtures containing high concentrations of other
materials. By employing a column
coupling procedure a large amount of sample can be used without the need for
sample pre-treatment such as extraction followed by concentration. The coupled column system that will be
described has a wide range of application as it can also be used to employ
other techniques in tandem with isotachophoresis such
as liquid chromatography or zone electrophoresis etc.
A diagram of a coupled column isotachophoresis
instrument is shown in figure 15. The coupled column apparatus consists of
three main parts, vis.,
1. A
pre-separation section
2. A
bifurcation section with a fitted detector
3. A
final separation section
In turn the bifurcation block has also three important
components, viz.,
a A
relatively wide bore hole (0.8 mm) in which the detector sensor(s) is mounted
and which is connected to the wide bore separation tube in the pre-separation
section.-
b
A narrow bore hole (0.2 mm) in line with the wide bore hole
(0.8 mm) to which the PTFE narrow bore
separation tube is connected.
C A rectangular hole (0.05mm x 1.0 mm) at right-angles to the other holes
which forms an electrical connection to the counter electrode compartment
applied during the pre-preparation.
![]()
![]()
![]()
![]()

Figure
15. A Coupled
Column Apparatus
A diagram of a commercially available isotachophoresis
instrument manufactured by Flowspek is shown in
figure 16.
The separation sections are in line with each other so that
zone mixing is minimised during trapping. To select the chosen band with
precision, the time interval between the detector sensing this band of interest
and the junction to the separation section can be calculated from the value of dV/dt of the constant current power supply in a leading
electrolyte/terminating electrolyte experiment. Thus, those zones requiring
further separation can be selected even if they are not migrating sequentially.

Figure
16. A Commercial Isotachophoresis Instrument
The choice of the bands of interest can be programmed so
that the selection is automatic and consequently highly reproducible. Using
this technique, the load that can be used for analysis can be increased by an
order of magnitude.